How Far Can You Run an Accelerometer Cable?

Maximum Cable Length for IEPE and Charge-Mode Accelerometers

An IEPE accelerometer can usually drive 30 m (100 ft) of cable with no special care, and hundreds of meters with the right supply current. A charge-mode accelerometer is different: keep the low-noise cable between the sensor and the charge amplifier as short as practical, with 10 m (33 ft) as a sensible upper limit.

Those two answers come from two different physical limits. This article explains each one, gives you the formula, and provides a table you can use to plan a cable run before you order.

What limits cable length on an IEPE accelerometer

On an IEPE sensor, the limit is cable capacitance, not noise. The built-in amplifier produces a low-impedance voltage signal, so a longer cable adds very little noise. What it does add is capacitance, typically around 100 pF per meter (30 pF per foot) for coaxial cable.

The sensor’s amplifier has to charge and discharge that capacitance on every cycle, and the only current it has to work with is the constant current from your signal conditioner, minus about 1 mA that the amplifier needs for itself. If the signal is large and fast, the amplifier runs out of current and cannot keep up.

The result is slew-rate distortion: the sine wave starts to look like a triangle wave, the measured amplitude drops, and false harmonics appear in the spectrum. Three things decide when this happens:

  • Total cable capacitance (length × capacitance per meter)
  • Peak signal voltage at the highest frequency of interest
  • Constant current supplied to the sensor

At low frequencies or small signal levels, cable length is rarely a concern at all.

The formula

The highest frequency an IEPE sensor can transmit without distortion over a given cable is:

fmax = 109 × (Ic − 1) / (2π × C × V)

Where fmax is in Hz, C is the total cable capacitance in pF, V is the peak signal voltage in volts, and Ic is the constant current in mA. The 1 mA subtracted from Ic is the current the sensor’s own electronics consume.

Worked example. A 30 m cable at 100 pF/m has 3,000 pF of capacitance. With a full ±5 V signal and a 4 mA supply, fmax is about 31.8 kHz, well above the range of most accelerometers. Drop the supply to 2 mA and fmax falls to about 10.6 kHz, which is now close to the working range of a general purpose sensor.

Note that V is the signal you actually expect, not the sensor’s full-scale output. A 100 mV/g accelerometer measuring 10 g peak produces 1 V, so it can drive five times the cable length of the same sensor at its full 50 g range.

Rearranged to solve for length, with Cm as the cable capacitance in pF per meter:

Lmax = 109 × (Ic − 1) / (2π × f × V × Cm)

Maximum IEPE cable length by current and frequency

The table shows the calculated cable length at which distortion begins, for a full ±5 V signal on 100 pF/m (30 pF/ft) cable.

Constant current1 kHz5 kHz10 kHz20 kHz
2 mA318 m (1,040 ft)64 m (210 ft)32 m (105 ft)16 m (52 ft)
4 mA955 m (3,130 ft)191 m (625 ft)95 m (313 ft)48 m (157 ft)
10 mA2,865 m (9,400 ft)573 m (1,880 ft)286 m (940 ft)143 m (470 ft)
20 mA6,048 m (19,800 ft)1,210 m (3,970 ft)605 m (1,980 ft)302 m (990 ft)

For planning, use half of these values. The formula marks the point where distortion starts, so a safety factor of 1.5 to 2 on frequency keeps the response flat across your whole range of interest.

Three adjustments to keep in mind:

  • Smaller signals go further. At ±1 V peak, every length in the table is five times longer.
  • Your cable may differ. Scale the lengths by 100 divided by your cable’s capacitance in pF/m.
  • Very long runs need a test. Beyond a few hundred meters, cable resistance and noise pickup also matter, so treat the largest values as theoretical and verify the installed system.

Practical tips for long IEPE runs

  • Raise the constant current. This is the most direct fix. Many battery-powered signal conditioners supply only 2 mA, which is the most limiting case. Line-powered conditioners and DAQ front ends often offer 4 mA to 20 mA. Check the sensor datasheet for its maximum current.
  • Do not over-drive short cables. Current the cable does not need turns into heat inside the sensor. Keep the current moderate on short runs and when the sensor is operating near its upper temperature limit.
  • Choose a lower capacitance cable. Halving the pF per meter doubles the usable length.
  • Match sensitivity to the signal. If the vibration level is high, a lower sensitivity sensor produces a smaller voltage swing and drives a long cable more easily.
  • Route and ground carefully. Use shielded cable, keep it away from power cables and variable frequency drives, and ground the shield at one end only to avoid ground loops.
  • Test the installed system. Compare a reading taken through the full cable against one taken through a short cable, at the highest frequency and amplitude you expect. If they match, the cable is not limiting you.

Charge-mode accelerometers: keep it short

On a charge-mode sensor, the limit is noise, not frequency response. The sensor has no built-in electronics, so the cable carries a tiny high-impedance charge signal all the way to the charge amplifier. Anything that happens to that cable happens to your measurement.

The good news is that with a charge amplifier, sensitivity in pC/g does not change with cable length. The concerns are elsewhere:

  • Amplifier noise rises with capacitance. The noise floor of a charge amplifier increases with the total capacitance at its input, which is the sensor plus the cable. Every added meter costs resolution.
  • Cable motion creates false signal. Flexing a cable generates charge (the triboelectric effect) that the amplifier cannot tell apart from the sensor output. Always use low-noise cable and clamp it close to the sensor.
  • A long high-impedance line is an antenna. It picks up electrical interference far more readily than an IEPE cable.
  • Contamination causes drift. Moisture or dirt in a connector lowers insulation resistance and degrades low-frequency response, and a longer cable usually means more connections.
  • The amplifier has a capacitance limit. Sensor capacitance plus cable capacitance must stay below the amplifier’s maximum source capacitance, or the signal is attenuated.

Recommended length: as short as practical, and no more than about 10 m (33 ft) of low-noise cable for general measurement work. Longer runs can work with high signal levels, but check the amplifier’s capacitance limit and expect a higher noise floor.

If you need distance, place an in-line charge converter a short distance from the sensor, outside the hot zone. The converter outputs an IEPE signal, so the rest of the run can use ordinary coaxial cable and follows the IEPE table above.

Quick reference

Sensor and cable runRecommendation
IEPE, up to 30 m (100 ft)No special care needed for typical vibration work up to 10 kHz
IEPE, 30 to 100 m (100 to 330 ft)Use 4 mA or more; run the formula if you need 5 kHz and above at high amplitude
IEPE, 100 to 300 m (330 to 1,000 ft)Use 10 to 20 mA; calculate with a 2x margin
IEPE, beyond 300 m (1,000 ft)Feasible at low frequency; calculate and test the installed system
Charge mode, sensor to amplifierAs short as practical, 10 m (33 ft) or less, low-noise cable only
Charge mode, long distanceIn-line charge converter near the sensor, then follow the IEPE rows

Planning a long cable run?

LNS Dynamics supplies IEPE and charge-mode accelerometers, low-noise and general-purpose cable assemblies, and signal conditioning, with cables built to the length your installation needs. Use the LNS cable builder to configure a cable, or contact us with your sensor model, cable length, and frequency range, and we will help you check the numbers.